Zetasizer µV - specification
The Zetasizer µV is dedicated to the measurement of the size and molecular weight of proteins, by both batch and SEC-LS modes.
Zetasizer News - changes in specification
System Specifications
General
- Temperature control range
- 2°C - 90°C +/- 0.1°C*
- Measurement angle
- 90°
- Condensation control
- Purge facility using dry air
- Standard laser
- 60mW, 830nm
- Correlator
- Min. sample time 480ns, max. 3600s. 264 channels
Options
- 21 CFR part 11 software
- Enables an operating mode that assists
with ER/ES compliance
System
- Compliance
- CE, Product: laser class 1, EN 60825-1:2001 and CDRH
- Dimensions, weight, power
- System 350mm x 410mm x 170mm (W,D,H), 14kg, 100W
- Minimum PC specification
- Pentium P4 PC 2.8GHz Dual core processor,
2GB RAM, 160GB hard disk drive, 1024 x 768
screen resolution, CD-ROM drive, USB port,
Windows XP Pro (SP3), Windows Vista or
Windows 7.
Note: Only the 32-bit Windows 7 and
Windows XP Pro operating systems
are supported.
Specifications for operation in cuvette mode
Size
- Parameters measured
- Hydrodynamic size, molecular weight
of Molecular solutions, e.g. proteins - Size range (radius)
- 0.15nm - 1 micron**
- Minimum sample volume
- 2μL
- Sensitivity
- 0.1mg/mL of 15kDa protein
Molecular weight
- Molecular weight range (estimated from hydrodynamic diameter)
- 42Da to >106Da†
- Molecular weight range (calculated using Debye plot)
- 980Da to >106Da†
- Minimum sample volume
- 2μL
- Software
- Results output
- Molecular size, molecular weight, A2, size distribution
Specifications for operation in chromatography detector mode
General
- Parameters measured
- Molecular weight, Molecular size
- Flow cell volume
- 8μL
Molecular weight
- Molecular weight range
- <500Da to >106Da†
- Software
- Results output
- Mn, Mw, Mz, Mp, Mw/Mn, Rh, Weight Fraction,
Concentration, dn/dc, dA/dc. A2 ††
OmniFACE interface specification
- Acquisition
- 2 channels RI, UV
- Dimensions, weight, power
- 260mm x 260mm x 60mm (WxDxH)
+ power adapter, max. 65W
Note:
* Temperature accuracy, 0.1°C at 25°C, 0.2°C at 2°C and 0.5°C at 90°C
**Peak mode range (radius), 0.3nm - 0.5 microns, sample dependent
† Sample dependent
††When used in conjunction with a concentration detector