System Specifications

General

Temperature control range
2°C - 90°C +/- 0.1°C*
Measurement angle
90°
Condensation control
Purge facility using dry air
Standard laser
60mW, 830nm
Correlator
Min. sample time 480ns, max. 3600s. 264 channels

Options

21 CFR part 11 software
Enables an operating mode that assists
with ER/ES compliance

System

Compliance
CE, Product: laser class 1, EN 60825-1:2001 and CDRH
Dimensions, weight, power
System 350mm x 410mm x 170mm (W,D,H), 14kg, 100W
Minimum PC specification
Pentium P4 PC 2.8GHz Dual core processor,
2GB RAM, 160GB hard disk drive, 1024 x 768
screen resolution, CD-ROM drive, USB port,
Windows XP Pro (SP3), Windows Vista or
Windows 7.
Note: Only the 32-bit Windows 7 and
Windows XP Pro operating systems
are supported.

Specifications for operation in cuvette mode

Size

Parameters measured
Hydrodynamic size, molecular weight
of Molecular solutions, e.g. proteins
Size range (radius)
0.15nm - 1 micron**
Minimum sample volume
2μL
Sensitivity
0.1mg/mL of 15kDa protein

Molecular weight

Molecular weight range (estimated from hydrodynamic diameter)
42Da to >106Da†
Molecular weight range (calculated using Debye plot)
980Da to >106Da†
Minimum sample volume
2μL
Software
Results output
Molecular size, molecular weight, A2, size distribution

Specifications for operation in chromatography detector mode

General

Parameters measured
Molecular weight, Molecular size
Flow cell volume
8μL

Molecular weight

Molecular weight range
<500Da to >106Da†
Software
Results output
Mn, Mw, Mz, Mp, Mw/Mn, Rh, Weight Fraction,
Concentration, dn/dc, dA/dc. A2 ††

OmniFACE interface specification

Acquisition
2 channels RI, UV
Dimensions, weight, power
260mm x 260mm x 60mm (WxDxH)
+ power adapter, max. 65W

Note:

* Temperature accuracy, 0.1°C at 25°C, 0.2°C at 2°C and 0.5°C at 90°C

**Peak mode range (radius), 0.3nm - 0.5 microns, sample dependent

† Sample dependent

††When used in conjunction with a concentration detector

 
 