Zetasizer µV
|
Zetasizer µV - specification
|
 |
 |
 |
|
 |
The Zetasizer µV is dedicated to the measurement of the size and molecular weight of proteins, by both batch and SEC-LS modes.
Zetasizer News - changes in specification
 |
| System Specifications |
 |
| General |
Temperature control range |
|
2°C - 90°C +/- 0.1°C* |
 |
| |
Measurement angle |
|
90° |
 |
| |
Condensation control |
|
Purge facility using dry air |
 |
| |
Standard laser |
|
60mW, 830nm |
 |
| |
Correlator |
|
Min. sample time 480ns, max. 3600s. 264 channels |
 |
| Options |
21 CFR part 11 software |
|
Enables an operating mode that assists with ER/ES compliance |
 |
| System |
Compliance |
|
CE, Product: laser class 1, EN 60825-1:2001 and CDRH |
 |
| |
Dimensions, weight, power |
|
System 350mm x 410mm x 170mm (W,D,H), 14kg, 100W |
 |
| |
Minimum PC specification |
|
Pentium P4 PC 2.8GHz Dual core processor, 2GB RAM, 160GB hard disk drive, 1024 x 768 screen resolution, CD-ROM drive, USB port, Windows XP Pro (SP3), Windows Vista or Windows 7.
Note: Only the 32-bit Windows 7 operating system is supported. |
 |
| Specifications for operation in cuvette mode |
 |
| |
Parameters measured |
|
Hydrodynamic size, molecular weight of Molecular solutions, e.g. proteins |
 |
| Size |
Size range (radius) |
|
0.15nm - 1 micron** |
 |
| |
Minimum sample volume |
|
2μL |
 |
| |
Sensitivity |
|
0.1mg/mL of 15kDa protein |
 |
| Molecular weight |
Molecular weight range (estimated from
hydrodynamic diameter) |
|
342Da to >106Da† |
 |
| |
Molecular weight range (calculated using
Debye plot) |
|
980Da to >106Da† |
 |
| |
Minimum sample volume |
|
2μL |
 |
| |
Software |
|
|
 |
| |
Results output |
|
Molecular size, molecular weight, A2, size distribution |
 |
| Specifications for operation in chromatography detector mode |
 |
| |
Parameters measured |
|
Molecular weight, Molecular size |
 |
| |
Flow cell volume |
|
8μL |
 |
| Molecular weight |
Molecular weight range |
|
<500Da to >106Da† |
 |
| |
Software |
|
|
 |
| |
Results output |
|
Mn, Mw, Mz, Mp, Mw/Mn, Rh, Weight Fraction, Concentration, dn/dc, dA/dc. A2 †† |
 |
| OmniFACE interface specification |
 |
| |
Acquisition |
|
2 channels RI, UV |
 |
| |
Dimensions, weight, power |
|
260mm x 260mm x 60mm (WxDxH) + power adapter, max. 65W |
 |
|
|
| |
Note:
* Temperature accuracy, 0.1°C at 25°C, 0.2°C at 2°C and 0.5°C at 90°C
**Peak mode range (radius), 0.3nm - 0.5 microns, sample dependent
† Sample dependent
††When used in conjunction with a concentration detector |